PN5300 Auto Injector


Environm. Temp.:
10-40°C
Humidity:
20-80% rel. Humidity
Dimensions:
300 x 575 x 360 mm
Weight:
21 kg
Power Requirements:
95-240 V, 50-60 Hz
Viscosity Range:
0.1 - 5 cP
Communication:
RS232C; option TCP/IP
Inputs/Outputs:
2 TTL in; 1 relay out
Sample Capacity:
2 Micro Titer plates according to SBS standards; 96-well high/low and 384-well low formats, 48-vial or 12-vial trays; any combination of plates is allowed, except for 384 low left and 96 high right side; standard 1.5 mL chromatography vials with/without inserts, 10 mL prep vials and small Eppendorf tubes can be used. Maximum vial/plate height is 47 mm overall. Automatic missing vial or well plate detection via internal sensor.
Loop Volume:
1-5000 µL program mable, 10 mL loop optional
Dispenser Syringe:
500 µL standard syringe, 2500 µL for prep option
Injection Valve:
Electrical switching time < 100 ms
Piercing Precision Needle:
± 0.6 mm
Wash Solvent:
Integrated wash solvent bottle
Wetted Parts:
SS316, PTFE, Tefzel, Vespel, Glass, Teflon, Peek; can vary with model metall-free etc.
Injection Cycle Time:
< 60 sec in all injection modes with injection volume < 100 µL including 300 µL wash.
Injection Modes:
Full loop, partial loopfill and µL pickup mode, pressure-assisted sample aspiration
Reproducibility:
RSD < 0.3 % for full loop injections (at 1.0 cP)
RSD < 0.5 % for partial loopfill injections (10 µL Vol.)
RSD < 1.0 % for µL pick-up injections (10 µL Vol.)
Sample Tray Cooling:
Built-in Peltier Range 4°C up to ambient -3°C
Sample Tray Cooling/Heating:
Built-in Peltier cooling/heating 4°-40°C
Prep Kit:
2.5 mL syringe; prep. valve; 10 mL sample loop; LSV needle; 24 pcs of 10 mL vials 2 prep sample trays, injection volume 1 µL up to 19.999 µL in 1 µL increments.