CF2000 Centrifugal FFF - Particle Separator

Environmental Conditions:
Relative humidity 20 – 80 %
(non-condensing) at an operating
temperature range of 10 – 30 °C
Measurement Range:
Nanoparticles: 10 nm - 20 µm
(Depending on sample material and density)
Minimum sizes achievable:
Al2O3, MnO2, Fe2O3, Clay, Si Au, Ag, are
12.5 nm, 12.5nm, 12.3 nm, 17.3 nm, 19.9 nm, 7.6 nm, 8.4 nm
Channel Cartridge System:
- Inner Volume: approx. 2.7 mL depending on spacer
- Inner Channel Thickness: 250 µm (others available)
- Channel Diameter: 200 mm
Applicable Solvent Systems:
All typical organic and aqueous solvents
Connected Detector Systems:
The following detection technologies can be used in FFF:
- Dynamic Light Scattering – DLS
- Static Light Scattering - MALS
- UV/VIS Absorbance
- Other Technologies: Please inquire on compatibility.
Flow Rate Ranges / Pressure:
- Main Flow 0.3 – 4.0 mL/min at channel outlet
- System Pressure up to 8 bar maximum
Centrifugal Force/Rotation Speed:
- Centrifugal acceleration 11 – 26369 m/sec²
(approx. 1.12 – 2688 g) - Maximum of 4900 rpm
- Constant, linear or exponential decay field
Required Analysis Time:
Typically 30 - 90 min; shorter/longer times possible
Injection Volumes:
Typically 10 – 100 µL
Injection Method:
Via manual injection valve or via PN5300 autosampler
Software System:
Unique NovaFFF software platform for complete system control, data acquisition, evaluation and report functions
Power Requirements:
100 – 240 VAC @ 50 – 60 Hz
PC Requirements:
Windows XP, min.1024 MB RAM, 1 RJ45 Ethernet LAN interface
Dimensions (DxWxH):
660 x 405 x 420 mm
Weight:
45 kg