Products >Modules >PN7000 Eluent Handling >PN7520 Eluent Degasser >Specifications

PN7520 Eluent Degasser

PN7520 Eluent Degasser

Degassing Principle:

By the use of an applied vacuum, dissolved gases are continuously removed through a semi-permeable Teflon® AF* membrane.

Flow Rate:

Max. 10 mL/min for each degassing channel

Channels:

2 channels
other channel numbers in request

Efficiency:

0.5 ppm oxygen at a flow rate of 0.5 mL/min

Wetted Materials:

Teflon® AF* (PTFE), PFA, PEEK, ETFE

Channel Volume:

0.45 mL per channel

Status Indicator:

Vacuum LED
Error LED
Power On LED

Error Message:

Contact closure relay
Capacity: 30 VDC, 1 A
Acoustic indicator

Environmental Conditions:

Operating Temperature Range:
5 – 40°C Relative Humidity: 20 – 80 % (non-condensing)

Power Requirements:

100 – 240 VAC @ 50 – 60 Hz, 100 W max.

Dimensions:

Width x Height x Length
270 x 90 x 430 mm

Weight:

5 kg

 

* Teflon AF® is a trademark of Dupont