Products >Modules >PN3000 Detectors >PN3150 RI Detector >Specifications

PN3150 RI Detector

PN3150 RI Detector

Environmental Conditions:

Relative humidity 20 – 80 % (non-condensing) at an operating temperature range of 10 – 30 °C

Detection Principle:

Deflection type differential RI detector

Refractive Index Range:

1.00 to 1.75

Temperature Settings:

30 °C to 50 °C in 1 °C steps

Flow Rates:

Micro cell: 0.1 – 1 mL/min (**)
Analytical cell: 1.0 – 5.0 mL/min (*)
Prep. cell: 50 – 200 mL/min (**)

Maximum Back Pressure:

50 kPa

Flow Cell Volume:

Micro: 2.5 µL (**)
Analytical: 8 µL (*)
Preparative: 8 µL (**)

Internal Volume:

Inlet port to flow cell: approx. 60 µL
Flow cell to outlet port: approx. 520 µL
Total internal volume: approx. 590 µL

Wetted Materials:

SST316, Teflon, Quartz Glass

Response Time:

0.1; 0.25; 0.5; 1.0; 1.5; 2; 3; 6 s

Measuring Range:

0.25 – 512 µRIU

Linearity:

600 µRIU

Noise Level:

2.5 nRIU (response: 1.5 seconds)

Integrator Output:

0 – 1 V

Recorder Output:

0 – 10 mV

Power Consumption:

150 VA (maximum)

Power Requirements:

100 – 240 VAC @ 50 – 60 Hz

Dimensions (DxWxH):

450 x 270 x 160 mm

Weight:

13 kg

 

(*) Standard, most sold
(**) On request