Environmental Conditions:
Relative humidity 20 – 80 % (non-condensing) at an operating temperature range of 10 – 30 °C
Detection Principle:
Deflection type differential RI detector
Refractive Index Range:
1.00 to 1.75
Temperature Settings:
30 °C to 50 °C in 1 °C steps
Flow Rates:
Micro cell: 0.1 – 1 mL/min (**)
Analytical cell: 1.0 – 5.0 mL/min (*)
Prep. cell: 50 – 200 mL/min (**)
Maximum Back Pressure:
50 kPa
Flow Cell Volume:
Micro: 2.5 µL (**)
Analytical: 8 µL (*)
Preparative: 8 µL (**)
Internal Volume:
Inlet port to flow cell: approx. 60 µL
Flow cell to outlet port: approx. 520 µL
Total internal volume: approx. 590 µL
Wetted Materials:
SST316, Teflon, Quartz Glass
Response Time:
0.1; 0.25; 0.5; 1.0; 1.5; 2; 3; 6 s
Measuring Range:
0.25 – 512 µRIU
Linearity:
600 µRIU
Noise Level:
2.5 nRIU (response: 1.5 seconds)
Integrator Output:
0 – 1 V
Recorder Output:
0 – 10 mV
Power Consumption:
150 VA (maximum)
Power Requirements:
100 – 240 VAC @ 50 – 60 Hz
Dimensions (DxWxH):
450 x 270 x 160 mm
Weight:
13 kg
(*) Standard, most sold
(**) On request